Gas Migration Analyzer Apparatus, Cement Hydration Analyzer                                          Model: TG-7200TG-7200 Gas Migration Analyzer to realistically simulate gas migration scenarios of varying severity.Model TG-7200 Gas Migration Analyzer is a precision instrument that measures four key aspects of oil-well cement:* its susceptibility to gas migration;* its degree of hydration;* its shrinkage during curing;* the gas permeability of the cement.Model TG-7200 is a closed sys...


  • Port: Dalian, Ningbo,Guangzhou
  • Payment Terms: L/C,D/A,D/P,T/T
  • Product Detail

    Product Tags

    Gas Migration Analyzer Apparatus, Cement Hydration Analyzer

                                              Model: TG-7200

    TG-7200 Gas Migration Analyzer to realistically simulate gas migration scenarios of varying severity.

    Model TG-7200 Gas Migration Analyzer is a precision instrument that measures four key aspects of oil-well cement:

    * its susceptibility to gas migration;

    * its degree of hydration;

    * its shrinkage during curing;

    * the gas permeability of the cement.

    Model TG-7200 is a closed system in which nitrogen gas is injected into the bottom of a cement slurry at any time during its hydration (setting). The cement’s susceptibility to gas migration is determined by whether or not the nitrogen gas injection pressure is transmitted up through the column of cement to the opposite (top) side of the cement sample where the pore pressure is measured.

    Gas migration will result in the pore pressure of the sample rising and possibly becoming and/or remaining equal to the gas injection pressure. If no gas migration occurs, the pore pressure will continue to drop (due to the shrinkage and loss of fluid communication through the sample) during hydration, possibly continuing to reach a vacuum.

    Model TG-7200 is designed to be as easy to use as possible. The instrument’s software controls, records and displays all test measurements in real time.

    Test conditions, including the static gel strength profile of the cement, are easily programmed and controlled through this system. As the system is designed to avoid line plugging, clean-up after testing is both quick and easy.

    Specifications

    Temperature Range   20 - 163ºC

    Pressures Maximum  1000 psi  

    Pressure Measurement Accuracy  0.2% of Full Scale

    Pressure Measurement Resolution  0.25 psi  

    Pressure Control Accuracy  ±10-20 psi  

    Cement Temperature & Cabinet Temperature

    Measurement Accuracy        ±0.5°C

    Measurement Resolution     0.1°C

    Oven Temperature

    Measurement Accuracy      ±1°C

    Measurement Resolution 0.1°C

    Temperature Control Stability ±0.5°

    Gas Injection Flow Rate 0 – 5 sccm of nitrogen

    Confining Flow Rate 0-5 sccm of nitrogen


     

    Flow Rate Measurement Accuracy

    1.2 % of Full Scale(±0.06 sccm)

    Flow Rate Measurement Resolution  0.01 sccm

    Cell Volumes

    Cement sample 417 cm3     Accumulator 100 cm3

    Utilities

    Water

    20-80 psi  Air

    20-160 psi  

    Nitrogen

    1000-3000 psi   

    Power Supply

    220VAC±10%,50/60 Hz 

    Dimensions

    90x110x75 cm  

     

     

     

    Specifications

    Temperature Range Ambientto325ºF(163ºC) Pressures Maximum 1000 psi / 6.9 MPa

    PressureMeasurementAccuracy

    0.2%ofFullScale

    PressureMeasurementResolution

    0.25psi/1.7kPa

    PressureControlAccuracy

    ±10-20psi/70-140kPa

    CementTemperatureandCabinetTemperature

    Measurement Accuracy    ±0.5°C Measurement Resolution0.1°C Oven TemperatureMeasurement Accuracy                           ±1°C Measurement Resolution0.1°C Temperature Control Stability

    ±0.5°

    GasInjectionFlowRate 0 – 5 sccm of nitrogen Confining Flow Rate

    0-5sccmofnitrogen


     

    FlowRateMeasurementAccuracy

    1.2 %ofFull Scale(±0.06sccm)

    FlowRateMeasurementResolution

    0.01 sccm

    CellVolumes

    Cementsample417cm3 Accumulator 100 cm3 Utilities

    Water

    20-80psi/140–550kPa Air

    20-160psi/140—1100kPa

    Nitrogen

    1000-3000psi/7-21MPa

    Power Supply

    220VAC±15%,50/60 Hz,10A 1-phase

    PhysicalDimensions

    Dimensions(hxwxd)

    92x112x74cm Weight

    1000lb/450kg

     